6.8
/ 10
MEDIUM
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Description
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software. Other related CVE's are CVE-2025-14096 and CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required configuration for Exposure:
Physical access to the analyzer is needed.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required configuration for Exposure:
Physical access to the analyzer is needed.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.
Basic Information
ID
CVE-2025-14095
Source
Radiometer
Published
Dec 17, 2025 at 11:45
Modified
Dec 17, 2025 at 11:54
Affected Product
Vendor
Radiometer Medical Aps
Product
ABL90 FLEX and ABL90 FLEX PLUS Analyzers
Version
All application software versions with Windows 7, Windows XP as underlying OS
Affected Versions
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers All application software versions with Windows 7, Windows XP as underlying OS
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 10 as underlying OS
Radiometer Medical Aps AQT90 FLEX Analyzers All Application software versions <= 8.13 MR2
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20 MR2 with Windows 10 as underlying OS
Radiometer Medical Aps ABL9 Analyzers Application software versions < 1.5.0
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 10 as underlying OS
Radiometer Medical Aps AQT90 FLEX Analyzers All Application software versions <= 8.13 MR2
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20 MR2 with Windows 10 as underlying OS
Radiometer Medical Aps ABL9 Analyzers Application software versions < 1.5.0