CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.
Other related CVE's are CVE-2025-14095 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure:
Attacker requires physical access to the analyzer.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Other related CVE's are CVE-2025-14095 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure:
Attacker requires physical access to the analyzer.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Basic Information
ID
CVE-2025-14096
Source
Radiometer
Published
Dec 17, 2025 at 12:19
Modified
Dec 17, 2025 at 13:03
Affected Product
Vendor
Radiometer Medical Aps
Product
ABL90 FLEX and ABL90 FLEX PLUS Analyzers
Version
Windows 7 Operating system
Affected Versions
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows 7 Operating system
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows XP Operating system
Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system
Radiometer Medical Aps AQT90 FLEX Analyzers Windows XP Operating system
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Operating system
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows XP Operating system
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows XP Operating system
Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system
Radiometer Medical Aps AQT90 FLEX Analyzers Windows XP Operating system
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Operating system