CVE 7.5 HIGH

Path Traversal Vulnerability in LabOne User Interface_CVE-2026-6903

7.5 / 10
HIGH
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

Description

The LabOne Web Server, backing the LabOne User Interface, contains insufficient input validation in its file access functionality. An unauthenticated attacker could exploit this vulnerability to read arbitrary files on the host system that are accessible to the operating system user running the LabOne software.

Additionally, the Web Server does not sufficiently restrict cross-origin requests, which could allow a remote attacker to trigger file access from a victim's browser by directing the victim to a malicious website.

The vulnerability is only exploitable when the LabOne Web Server is running. Installations using only the LabOne APIs without starting the Web Server are not exposed.

Basic Information

ID CVE-2026-6903
Source NCSC.ch
Published Apr 23, 2026 at 09:45

Affected Product

Vendor Zurich Instruments
Product LabOne
Affected Versions Zurich Instruments LabOne 0